QLAB Analysis For
3D Volumes (IE 33 Philips)
>> Select Image
Use pointer to
select image from thumbnails on the right, double click on image.
>> Start QLAB
Select “QLAB”
from left touch screen, select “3DQ Adv” from left touch screen.
>> Adjust planes
4 images appear
on screen
Go to upper left
image, grab intersection of blue and red lines, center at MV annulus.
Go to upper right
image, grab intersection of blue and green lines, center at MV annulus.
Go to upper
images and tilt red and green line through apex.
Go to lower left
images and tilt green and red line to optimize upper 2 image.
Raise blue line
in either of upper images until RV is well visualized in short axis view.
Go to lower left
image, point yellow arrow toward middle of ventricular septum.
-
Readjust any of the planes at anytime.
-
Use “slice plane views”, lower right image, to doubled
check plane placement.
>> Place End Diastolic Reference Points
Find end
diastolic frame.
Click on “ED Ref.
Points” in upper left corner of the screen.
“S” will appear,
place at septal/MV annulus point in upper left image.
“L” will appear,
place at lateral/MV annulus point in upper left image.
“A” will appear,
place at anterior/MV annulus point in upper right image.
“I” will appear,
place at inferior/MV annulus point in upper right image.
“Ap” will appear,
place at apical point in either of the upper images (where apex is best seen).
Use lower right image to help place apical point.
>> Edit Image
Tracking can be
doubled checked by moving blue plane up and down through ventricle, watch the
lower left image (short axis) to check tracking.
If image did not
track properly select “Add Ref. Points” from upper left side of screen.
Place points
where you think the endocardial border is, select “update image” from upper
left side of screen, border will move to these points.
Repeat if
necessary, select “Finish Edit” from upper left side of screen when all editing
is done.
>> Place End Systolic Reference Points
Find end systolic
frame by clicking on numbered boxes at bottom of screen or use arrows on
keyboard.
Readjust planes
if necessary.
Click on “ES Ref.
Points” in upper left corner of the screen.
“S” will appear,
place at septal/MV annulus point in upper left image.
“L” will appear,
place at lateral/MV annulus point in upper left image.
“A” will appear,
place at anterior/MV annulus point in upper right image.
“I” will appear,
place at inferior/MV annulus point in upper right image.
“Ap” will appear,
place at apical point in either of the upper images (where apex is best seen).
Use lower right image to help place apical point.
Image will track
endocardial border.
>> Edit Image
Tracking can be
doubled checked by moving blue plane up and down through ventricle, watch the
lower left image (short axis) to check tracking.
If image did not
track properly select “Add Ref. Points” from upper left side of screen.
Place points
where you think the endocardial border is, select “update image” from upper
left side of screen, border will move to these points.
Repeat if
necessary, select “Finish Edit” from upper left side of screen when all editing
is done.
>> Sequence Analysis
Select “Sequence
Analysis” from left side of screen, all frames will be tracked.
>> Results
· Press space bar
to show moving image.
· Select “Show
Segments” and “Show Ref. Mesh”, select “Hide Wave Forms”.
· Select “3D”” from
tool bar, rotate image and acquire.
· Select “i-slice
view” on left side of screen, magnify and charge to 4X4 view and acquire.
· Select “Show
Report Page” on left side of screen and acquire loop.